CONCENTRATE OF MOLECULAR IODINE DETECTION WITH PORTABLE COLORIMETER ON POLYURETHANE FOAM

UDC 543.432'831

Authors

  • A. Тrohimenko, Cand. Sci. (Chem.) Taras Shevchenko National University of Kyiv image/svg+xml
  • O. Zaporozhets, DSc (Chemistry) Taras Shevchenko National University of Kyiv image/svg+xml

Keywords:

iodine, iodide, polyurethane foam, portable photometer, solid-phase spectrophotometry

Abstract

The possibility to use the portable photometer for registering color characteristics of the polyurethane foam (PUF) was studied. For the system I2-PUF was plotted calibration curve: dependence of the intensity of the B color channel from the amount of iodine in the sample. Construction of calibration dependences was performed in the following way. In syringes was taken from 0,25 to 10,0 cm3 of a standard solution of iodine with a concentration of 1,2 mg/dm3 (pH~2) and sulfate acid (pH~2) to a total volume of 10,0 cm3, stirred, removed the iodine on surface of PUF, and recorded the analytical signal at 370 nm with two method relative to pure tablets sorbent. To build a calibration curve with iodide carried out a preliminary oxidation of iodide to I2 using oxone. The dependence of the analytical signal of iodine on the surface of the sorbent using a portable photometer described by the equation y=1,32+0,92.10-4.с (mg/dm3), R2=0,998, where y – the analytical signal B-channel. The results for iodine PUF and colorimetric methods solid-phase spectrophotometry almost the same. Range of linearity calibration curve is wider in the case of colorimetric detection. The results of determination supplementation of I2 and I- to bottled drinking water using a portable photometer. Established that confidence interval changes from the maximum value for ±9,7 % с(І2)=0,1 mg/dm3 to a minimum for ±6,0 % с(І2)=1,0 mg/dm3 (n=5, P=0,95). Portable colorimeter due to its sensitivity and accuracy promising and convenient for use in situ (at sampling site).

References

1. Ivanov V.M., Monogarova O.V., Oskolok K.V. J. Anal. Chem., 2015, 70(10), 1165–1178.

2. Khimchenko S.V., Eksperiandova L.P. Color Measurements in Instrumental and Visual Test Analysis. Lambert Academic Publishing, 2014, 220 p. (In Russian)

3. Ivanov V.M., Kuznetsova О. V. Russ. Chem. Rev., 2001, 70(5), 357–374.

4. Apyari V.V., Dmitrienko S.G. J. Anal. Chem., 2008, 63(6), 530–537.

5. Trokhimenko A. Yu., Zaporozhets O. A. J. Anal. Chem., 2014, 69(5), 408–412.

Published

2016-12-06

How to Cite

CONCENTRATE OF MOLECULAR IODINE DETECTION WITH PORTABLE COLORIMETER ON POLYURETHANE FOAM: UDC 543.432’831. (2016). Bulletin of the Taras Shevchenko National University of Kyiv. Chemistry, 52(1), 36-38. https://chemistry.bulletin.knu.ua/article/view/8590

Most read articles by the same author(s)

<< < 1 2